HL-2000-CAL

The HL-2000-CAL Calibrated Tungsten Halogen Light Source can calibrate the absolute spectral response of a radiometric system. With this NIST-traceable light source and SpectraSuite, you can determine known absolute intensity values at wavelengths from 300-1050 nm. The HL-2000-CAL Calibrated Tungsten Halogen Light Source can be used to calibrate the absolute spectral response of a radiometric system. With this NIST-traceable light source and SpectraSuite Software.

 

 

Specifications

 
Spectral range (calibrated): 300-1050 nm (calibrated)
Power consumption:

600 mA @ 12 VDC

Power output: 6.5 watts
Bulb life:

900 hours (recommend recalibration after 50 hours of use)

Recalibration: Required after ~50 hours of operation
Bulb color temperature:

3100 K

Output to bulb: 5 volts/1.3 amps
Output regulation: 0.2% voltage
Time to stabilized output: ~30 minutes
Connector: SMA 905 for fiber; 6.35-mm barrel for cosine corrector; PTFE plug for integrating sphere